Input/output circuit
Fig. 2:
= input/output circuit cable

A = view A
Note:
The table below contains the values which depend on the type code (type of device).
Always remember to compare the type code in the table with the code on the name
plate of your device.
Transmitter Prosonic Flow 93***-***********A
Terminals
20
+
Designation
Functional
values
Intrinsically
safe circuit
U
=
m
I
=
m
Transmitter Prosonic Flow 93***-***********B
Terminals
20
+
Designation
Relay

Functional
60 V DC/100 mA
values
or
30 V AC/500 mA
Intrinsically
no
safe circuit
U
=
260 V AC
m
I
=
m
21
22
23
+
active: 24 V DC/25 mA
21
22
23
+
Relay

60 V DC/100 mA
or
30 V AC/500 mA
no
260 V AC
500 mA
24
25
26
+
+
Pulse/frequency
Current output
output

f = 2...10000 Hz
active: 0/4...20 mA
(max. 250 mA/20 ms),
R
L
>100 Ω
R
passive: max. 30 V DC
L
passive: 30 V DC/
250 mA
no
260 V AC
500 mA
24
25
26
+
+
Pulse/frequency
Current output
output

f = 2...10000 Hz
active: 0/4...20 mA
active: 24 V DC/25 mA
(max. 250 mA/20 ms),
R
L
>100 Ω
R
passive: max. 30 VDC
L
passive: 30 V DC/
250 mA
no
260 V AC

27
HART

< 700 Ω
R
L
HART ≥ 250 Ω
no
260 V AC
27
HART

< 700 Ω
R
L
HART ≥ 250 Ω
no
260 V AC
7

Publicité

Table des Matières
loading

Table des Matières