I/O Board wiring labels
I/O Board Part Number
P/N 191910
SENSOR INPUT(2)
TBxA - SENSOR 1
TBxB - SENSOR 2
Ch
ECOND
TB
CCOND
1
RCV+
2
RCV–
3
4
5
XMT+
XMT
6
XMT–
1,2
7
RCV
8
R-SHLD
9
TEMP+
TEMP+
10
TEMP–
TEMP–
11
X-SHLD
SHIELD
12
13-18
I/O Board Part Number
P/N 191910
SENSOR INPUT(2)
TBxA - SENSOR 1
TBxB - SENSOR 2
TB
Ch
ECOND
CCOND
1
RCV+
2
RCV–
3
4
5
XMT+
XMT
6
XMT–
1,2
7
RCV
8
R-SHLD
9
TEMP+
TEMP+
10
TEMP–
TEMP–
11
X-SHLD
SHIELD
12
13-18
Figure 7 Câblage de Carte de Capteur Double, Numéro de Pièce 191910 - Conductivité
I/O Board P/N
Identification Labels
Channel 1
pH/ORP DIS
1
2
3
+5V
4
–5V
XMT
RED
5
6
RCV BLACK
IN+
7
IN–
8
TEMP+ GRN
TEMP+
9
TEMP– WHT
TEMP–
10
SHIELD
SHIELD
11
12
Channel 2
pH/ORP DIS
RCV + RED
1
RCV – BLK
2
3
+5V
4
–5V
XMT + WHT
5
XMT – BLK
6
7
IN+
R-SHLD
8
IN–
TEMP + GRN
9
TEMP+
TEMP – BLK
10
TEMP–
X-SHLD
11
SHIELD
12
Notes:
Identify P/N 191910 I/O Board and connect wires to the terminal blocks directly
18
below the I/O slot that the board is in.
Use the wiring label located on the front panel that has a matching I/O part number.
Either channel can support any of the sensor types listed.
I/O Boards 1 through 4
I/O Board 1-4
Terminal Blocks
Contacting Conductivity (CCOND)
Contacting Conductivity (CCOND)
(Wiring is typical of all three sensor options)
Conductivity
Electrode
High Pressure
GRN WHT RED BLK
Electrodeless Conductivity (ECOND)
24
General Purpose

Publicité