In the Configuration Menu the user can customize the device after his favor. The parameters are organized
in different levels. The following information and parameters are stored in there.
• Software version
• Mainboard serial number

• Display illumination

• Operating hours MD 11
• Operating hours Motor
• Operating hours of irrigation pump
• Activation of available handpiece
Be cautious when changing parameters. Unusual behavior of instruments while operating may provoke
false reactions and jeopardize the patient.
Every setting and the new behavior of the instrument has to be verified and tested.
Access to Configuration Menu:
• Press "HP" and "Speed ▼" key for about 3 sec. until you hear a tone.
On the display the first position of the configuration menu shows up.
The arrows at the beginning of the lines indicate
the operation in the Configuration Menu.
• Press "HP" or "PR" choose the desired parameter.
• Press "Speed ▼" to go to change mode.
• To change the value press "HP" for up or "PR" for down.
• To confirm the changes made, press "Speed ▼" for about one sec., until you hear a short tone.
• To abort the settings made press "Speed ▼" briefly, the setting returns to its previous value.
Exit from Configuration Menu:
• To leave the Configuration Menu press "HP" and "PR" for 3 sec., until you hear the long tone.

Parameter Level 1

Group/Parameter
Software/Version
Hardware/Serial number MB
Backlight/brightness (0. .10)
Operating hours/Control Unit
Operating hours/Motor
Operating hours/Pump
Error memory/ 1 – 8
16
Rights
Factory
read
VX.XXXX
read
XXXXXXXXXXX Shows serial number of the main board
read/write
9
read
0
read
0
read
0
read
0
• Speed limitation for each handpiece
• Range of action of AS mode
• Pump behavior
• Motor behavior
• Behavior of torque at reverse mode
• Reset to factory default
Definition
Shows current software version
Display brightness, changeable: 0, ... , 10
Shows operating hours of Control Unit
Shows operating hours of motor 1
Shows operating hours of pump
8 Error messages in chronological order.
MD 11 Operating Instructions, Art. No. 31665, V10/14