If the evaluation unit detects that the baby has not taken a breath for longer than about 20 seconds, the
following type of alarm goes off: fi rstly, there is a short acoustic pre-alarm followed by intensive acous-
tic alarm accompanied with the fl ickering red indicator. If the child is breathing too slowly – less than 8
intakes per minute, the following type of alarm goes off: there is an immediate acoustic alarm and the
green light is on constantly (it shows that there is a breathing activity) and the red light is fl ickering.
Check the baby. If she isn't breathing, try to wake her up. If she does not wake up, start administering fi rst
aid immediately (release her airways, start mouth-to-mouth resuscitation, etc.). We recommend contacting
a doctor depending on the circumstances. In some cases the alarm signal itself rouses the baby strongly
enough to make her take a breath. To turn the alarm off, press the switch on the evaluation unit. Exceptio-
nally, a false alarm may occur, which is the case especially when the sensor pad is not fi tted properly (see
the Installation Instructions and the Frequently Asked Questions) or when the baby can crawl and moves
outside the reach of the sensor pad inside the crib.
We hope you hear the alarm signal only once: when testing our product.
A sensor pad can be bought as a separate item. It is available under item No. BM-02D.
It is extremely handy when the device is used in several places; e.g. in another crib, at grandma's, etc.
In this case the evaluation unit is the only thing you have to remember to take along.
Check the operation of the device as follows (recommended testing frequency is once a week):
1. When the baby is in the crib and the Nanny is on, the green indicator must fl icker with the rhythm of the
baby's breath.
2. Leave the Nanny on and take the baby out of the crib. The indicator usually keeps fl ickering for a while
until the crib stops shaking. Do not hold on to the crib because the Nanny could detect your breath and
movements.
3. When the crib has been still for about 20 seconds the evaluation unit gives off a warning beep and then an
alarm signal is activated. While the alarm is on, the red indicator keeps fl ickering. The alarm can be turned
off with the switch.
If the Nanny does not work check whether:
1. The evaluation unit confi rms the activation of the Nanny by a low beep and fl ickering. If it does not, check
the batteries.
2. The green indicator stops fl ickering after taking the baby out of the crib. If it does not, the Nanny detects
disturbing vibrations – see the following information.
Important information:
The device uses a highly sensitive sensor to detect breathing. Its operation can be affected by shakes of
the crib, fl oor or the whole building. This is why the crib must not touch a bed in which somebody is sle-
16
MFV52110
EN
ALARM STATUS
SENSOR PAD
FUNCTION TEST