1. Replace the lancing device cap with the clear cap. (o)
2. Pull the cocking control out until the orange bar appears on
the release button window.
Important!
• Choose a different spot each time you test. Repeated
punctures at the same spot may cause soreness and calluses.
• Avoid lancing the areas with obvious veins to avoid excessive
bleeding.
• It is recommended to discard the first drop of blood as it might
contain tissue fluid, which may affect the test result.
Make Blood Glucose Test
1. Insert the test strip into the test slot of the device. Wait for the
device to display the blood drop " ". (e)
2. Select the appropriate measuring mode by pressing MAIN.
• General Tests - any time of day without regard to time since
the last meal.
• AC (
) - no food intake for at least 8 hours.
• PC (
) - 2 hours after a meal.
• QC (
) - testing with control solution.
3. Obtain a blood sample. (m)
Use the pre-set lancing device to puncture your desired site.
After penetration, discard the first drop of blood with a clean
tissue or cotton. Gently squeeze the punctured area to obtain
another drop of blood. Be careful NOT to smear the blood
sample. The volume of blood sample must be at least 0.5
microliter (μL) for blood glucose testing.
4. Apply the blood sample. (p)
Move your finger to meet the absorbent hole of the test strip
and the drop will be automatically be drawn onto the test strip.
Remove your finger until the confirmation window is filled. The
meter begins to count down. Do not remove your finger until
you hear a beep sound.
5. Read your result. (q)
The results of your blood glucose test will appear after the
meter counts down to 0. The results will be stored automatically
in the meter memory.
Announcement of Test Result by Universal Tone
The blood glucose results will be broken down into individual
digits and each digit represents the corresponding number of
beeps.
The result is announced three times in succession and each time
is preceded by two quick beeps. So you will hear: 2 quick beeps –
results – 2 quick beeps – results – 2 quick beeps – result.
For mg/dL meters, the hundreds are always announced, even
when the result is below 100.
Examples:
80 mg/dL is announced as 1 long beep (0) – 1 single pause – 8
single beeps (8) – 1 single pause – 1 long beep (0)
182 mg/dL is announced as 1 single beep (1) – 1 single pause – 8
single beeps (8) – 1 single pause – 2 single beeps (2)
For mmol/L meters, the tens are always announced, even when
EN-7