cleaning is necessary. To decrease the cleaning times, increase the
temperature or use an ultrasonic bath. To complete the cleaning, rinse
a few times with deionized water and then let the sample cell air dry.
Sample cells may also be cleaned with acid, followed by a thorough
rinse with deionized water.
Note: Always use acid to clean sample cells that were used for low-level metal
tests.
Special cleaning methods are necessary for individual procedures.
When a brush is used to clean sample cells, take extra care to avoid
scratches on the interior surfaces of the sample cells.
9.3 Replace the batteries
Replace the batteries when the battery power level is low. Refer to
Install the batteries
Section 10 Troubleshooting
Error
E-00
E-01
E-02
7
When an E-01 or E-02 error occurs on a measurement, the
display shows "_.__". The decimal place depends on the
chemistry. If the E-01 or E-02 error occurs while the instrument is
set to zero, set the instrument to zero again.
on page 7.
Description
No Zero
Ambient light
7
error
7
LED error
Solution
In user calibration mode, a standard
solution was measured before the
instrument zero was set. Measure a blank
solution to set the instrument to zero.
There is ambient light in the cell holder.
Make sure that the instrument cap is fully
installed on the cell holder. Refer to
test
on page 11.
The LED (light source) is out of
regulation. Replace the batteries. Make
sure that the LED in the cell holder
comes on when
Do a
or
is pushed.