Hach DR300 Manuel De L'utilisateur page 15

Masquer les pouces Voir aussi pour DR300:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 51
Section 7 Show measurements
Note: The instrument saves a maximum of 50 measurements. After
50 measurements are done, new measurements replace the oldest
measurements.
1. Push and hold
2. Push
until "rCL" (recall) shows, then push
"– 01 –" shows. Measurement 01 is the last measurement done.
3. Push
to scroll forward.
The measurement number is followed by the measurement value
and then the time.
4. To go to a measurement number, push
number shows, then push
Note: Measurements cannot be deleted.
5. Push and hold
Section 8 Calibration
This instrument is calibrated at the factory. No user calibration is
necessary.
8.1 Standard calibration adjust
Use the standard calibration adjust (SCA) option when a calibration
must be adjusted to meet regulatory requirements. The factory
calibration is adjusted slightly with the standard calibration adjust
(SCA) option so that the instrument shows the expected value of the
standard solution. The adjusted calibration is then used for all test
results. This adjustment can increase the test accuracy when there are
slight variations in the reagents or instruments.
Note: For instruments with factory-calibrated ranges or methods, the standard
calibration adjust (SCA) feature is disabled when a user-entered calibration is
entered into the instrument. To set SCA back to on, set the instrument to the
factory default calibration. Refer to
on page 20.
for 3 seconds.
or
for 3 seconds to go back to measurement mode.
Set to the factory default calibration
.
until a measurement
.
English 15