Insert ceiling tube 1 into ceiling bracket 2 and
drill with d= 9mm 3. Use the existing hole in the
ceiling bearing as a guide.
Make 4 opposited drill hole markings (Ø13mm),
note the position of the Ø60mm drill hole for
electrical connections
Check clearances
DANGER
Risk of injury from falling parts
The ceiling bracket must be secured to the
fastening material, which must be suitable for
the corresponding ceiling condition.
DANGER
Risk of injury from falling parts
Mounting must be performed by two people.
CAUTION
Use protective equipment in line with the tool
manufacturer's instructions
Drill the bore holes, then blow them out using
bellows
20
DANGER
Risk of injury from falling light
The ceiling must be made of solid concrete to
guarantee a secure hold.
Use suitable ceiling anchors.
Disassemble the securing element from the
ceiling tube (Allen wrench is enclosed; do not
lose the threated pin!)
Insert the cable (with loose ends first) and pull
up to the stop
Pull the light cable of the ceiling tube through
the light mount.
Insert ceiling tube into light mount
Secure using the M8 safety screw and nut 1
Hold the ceiling bracket against the ceiling and
strike the ceiling anchor with a hammer
Tighten fastening in line with manufacturer's
data