SAFETY INSTRUCTIONS

1.1

Intended use

med®
The light D
VISIANO is an examination light.
Its intended use is to illuminate the body of the
patient
to
support
An interruption of the diagnosis or treatment
caused by a light failure is always possible without
posing any risk for the patient. The light cannot be
used in operating rooms.
1.2

User profiles

Medical professional
A medical professional is any person who has
completed medical training and is working in his or
her professional field.
Cleaning professional
A cleaning professional is trained in national and
job-related hygiene regulations.
Electrician
An electrician is trained in the fields of electronics
and electrical engineering and is familiar with the
relevant standards and regulations.
Qualified professional
A qualified professional is capable of mounting and
dismounting the light thanks to professional
training, knowledge and experience and knowledge
of the regulations.
1.3
Safety instructions
Operation by a medical professional
The instructions form part of the product.
They must be stored and made accessible to
all subsequent users.
Any work on the light (including repairs) must
be carried out by qualified electricians only.
Mounting must be performed by a qualified
professional only.
The light must not be altered or manipulated in
any way. Only approved original parts must be
used. Any use other than the intended use with
original parts may give rise to other technical
values and life-threatening danger.
Operation in potentially explosive areas is
prohibited. The light power supply is a
potential ignition source.
The light must only be operated in dry, dust-
free rooms.
The light must not be left switched on without
supervision.
For lights in protection class I, it is essential for
the earthing conductor to be connected to the
light housing.
Do not use any light that is damaged.
Defective cables are also potential hazards.
Do not place the cable close to any heat source
or on sharp edges.
Damage to the eyes. Never look directly into
the light source.
Replace any glass that is damaged before
operating the light again.
In order to prevent electric shocks, only
connect the light to a power supply with an
earth conductor (only C-version).
Luminaires which are equipped with a hospital
grade plug (UL 817, only US- and Canada
version), the earth connector is used as a
functional earth within the luminaire.
18
diagnosis
or
treatment.
Never place additional loads on the light head
or the arm.
The light must not be covered by a cloth or any
similar item while it is in operation.
The ventilation openings (where they exist)
must be free whenever the light is in
operation.
The light must not be operated near to external
heat sources that exceed the maximum
ambient temperature of the light.
The light must not be used in environmental
conditions other than those for which it is
intended.
The light must only be used for the intended
use described in this document.
The manufacturer cannot be held responsible
for any injury or damage that is caused as a
result of any use other than the intended use
or of any failure to comply with safety
instructions and warnings.
1.4

Warning levels

DANGER
Indication of hazards that can lead to death or
serious injury if measures are disregarded.
WARNING
Indication of hazards that can lead to injury if
measures are disregarded.
CAUTION
Indication of hazards that can lead to damage to
property if measures are disregarded.
1.5

Specific mounting instructions

VISIANO 20-2 C T1
Mounting material is not included.
The ceiling mounting must only be fixed to
ceilings that have a concrete stability class B25
(C20/25) or higher.
Solid ceiling reinforcement parts must not
come into contact with the ceiling mounting. If
there is any doubt, an authorized professional
must confirm that mounting is possible on the
mounting base being used.
Drilling must be done by specialists in
compliance with the drilling tolerances for the
reinforcement anchor that have been approved
by the manufacturer.
Screws must be tightened carefully using a
torque key that complies with the instructions
of the reinforcement anchor manufacturer.
VISIANO 20-2 P TX
Mounting material is not included.
When using the accessory "wall mount", the
installation
must
professional.
The wall must guarantee a firm hold.
Use only mounting material which is suitable
for the corresponding substrate.
be
undertaken
by
a