Always follow the instructions below in case a problem occurs.
Using methods other than those described may cause the pump to function incorrectly.
Condition
NO SET
• The set is not fitted or wrongly
• The pressure sensor area is dirty
PUSH STRT
• The pump has been untouched
END OF DOSE
• The programmed dose/volume
PROG
• No flow rate is installed.
DOOR
• The door is not correctly closed.
AIR
• The pump detected air in the
• The air sensor area is dirty.
• The feeding set is not inserted
BATT
• The battery capacity is too low.
the pump
remains working
Battery, "E" and "F"
• Battery failure.
flashing
FILL SET
• The pump is priming the set.
OCC IN
• The pump detected an upstream
• The pressure sensor area is dirty
OCC OUT
• The pump detected a
• The pressure sensor area is dirty
ZMNL900908_AW.indd 28
UK

PROBLEM SOLVER CHART

Cause
fitted in the pump.
(Fig 1, position 5 & 6).
in hold mode for 3 minutes or
more.
has been administered.
Flow rate = 0 ml/h
pump set.
correctly (Fig 1, position 8).
The pump is not able to deliver
highly viscous fluids at high flow
rates with the present charge
level of the battery.
occlusion between the pump
and the feeding bag.
(Fig 1, position 5 & 6).
downstream occlusion between
the pump and the patient.
(Fig 1, position 5 & 6).
Correction
• Stop the alarm by pressing "START/STOP"
• Insert the Flocare
Infinity
®
as indicated on the blister packaging (or refer to Fig
5a - 5d) and close the door.
• Restart the pump.
• Clean the sensors, reinsert the feeding set in the
pump and restart the pump.
• Stop the alarm and prolong the hold mode with
another 3 minutes by pressing "START/STOP" key.
• Program the pump and start it by pressing the
"START/STOP".
• Turn the pump off by pressing the
it during 2 seconds or
• Clear the memory of the total volume administered
(see paragraph: clearing the memory), reprogram a
new feeding schedule and start the pump by pressing
"START/STOP".
• Make sure the correct flow rate is programmed.
• Make sure the door is properly closed prior to
starting a feeding program.
• Make sure there is no excessive foam in the reservoir
and ensure there is enough feed in the reservoir for the
set therapy.
• If the feeding reservoir is empty, replace the reservoir and
continue feeding. If necessary, prime the set (FILL SET).
• Make sure the air sensor is clean.
• Make sure the feeding set is properly inserted in the
pump. (Fig 5a - 5d).
• Connect the charger to the pump and mains and
charge the pump for approx. 6 hours. During char-
ging the pump can be used.
• Turn the pump "off", contact your healthcare provider
or refer the pump to service.
• Press the "FILL SET" key another time to stop the
pump and bring it back in the hold mode.
• Stop the alarm by pressing the "START/STOP" key.
• Remove the feeding set out of the pump and check
the permeability by flushing the line.
• Re-insert the feeding set in the pump and restart.
• Clean the sensors, reinsert the feeding set in the
pump and restart the pump.
• Stop the alarm by pressing the "START/STOP" key.
• Remove the feeding set out of the pump and check
the permeability by flushing the line.
• Check the permeability of the feeding tube.
• Re-insert the feeding set in the pump and restart.
• Clean the sensors, reinsert the feeding set in the
pump and restart the pump.
-26-