The Flocare
Infinity
®
below. The customer or the user of the Flocare
such an environment.
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Proximity fields
from RF
wireless
communications
equipment IEC
61000-4-3
Radiated RF
RTCA/DO-160E
Section 20
NOTE 1
At 80 Mhz and 800 MHz, the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorp-
tion and reflection from structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted the-
oretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an elec-
tromagnetic site survey should be considered. If the measurement field strength in the location in which the
Flocare
Infinity
is used exceeds the applicable RF compliance level above, the Flocare
®
be observed to verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating the Flocare
b. Over the frequency range 150 KHz to 80 MHz, field strengths should be less than 3 V/m.
ZMNL900908_AW.indd 34
Guidance and manufacturer's Declaration – Electromagnetic immunity
enteral feeding pump is intended for use in the electromagnetic environment specified
®
IEC 60601 test level
Compliance level
3 Vrms
3 Vrms
150 KHz TO 80
MHz outside ISM
and amateur
radio bands
80% AM at 1 kHz
6 Vrms
6 Vrms
150 kHz to 80 MHz
in ISM and amateur
radio bands
80% AM at 1 kHz
10 V/m
10 V/m
80 MHz to 2.7 GHz
80% AM at 1 kHz
9 V/m to 28 V/m
9 V/m to 28 V/m
15 specific frequen-
cies up to 5,785 GHz
75 V/m
No Equipment
100 MHz to 8 GHz
Category specified
(75 V/m)
Infinity
enteral feeding pump should assure that it is used in
Electromagnetic environment – guidance
Portable and mobile RF communications
equipment should be used no closer to
any part of the Flocare
cables, than the recommended separation
distance calculated from the equation appli-
cable toe the frequency of the transmitter:
Recommended separation distance
d = 1.2CD P
d = 1.2CD P
d = 2.3CD P
Where P is the maximum output power
rating of the transmitter in watts (W) accor-
ding to the transmitter manufacturer and d
is the recommended separation distance
in metres (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey,
should be less than the compliance
a
level in each frequency range.
Interference may occur
in the vicinity of equipment
marked with the following
symbol:
Infinity
.
®
-32-
®
Infinity
, including
80 MHz to 800 MHz
800 MHz to 6 GHz
b
Infinity
should
®
4.05.21 16:31