The following information can be found in the display:
• Flow rate (ml/h), volume (ml), is displayed through the large characters.
Words below describe what the number relates to (rate, dose or volume).
The pump also displays messages, for example "end of dose" will appear when the pump has
finished delivering a single feed dose.
TOTAL VOLUME DEL
INTERVAL BOLUS
Figure 6
The back light of the display turns off 10 seconds after the last key is pressed.
The back light will switch on for 10 seconds when the charger is connected to the mains.
• The Flocare
Infinity
®
• Insert the pump set in the pump, switch the pump on and fill the pump set when necessary (see
paragraph: filling the pump set).
• The display shows the total volume delivered since the pump was last cleared.
• If necessary clear the volume delivered by pressing the "EFFACER" key.
• The last installed flow rate (ml/h) is now displayed.
• The pump is now in the "hold" mode and is ready for programming.
• We strongly recommend that for those patients where interruptions or delay in therapy could affect their health status, the
DOSE setting of the Infinity pump must be used and programmed.
d
DEBIT
d
• Adjust the flow rate (ml/h) if required by using the "+" or "-" key. Hold either key down to change
rapidly.
• If another parameter is indicated in the LCD, it will be necessary to first press the "ml/h" key followed
by the "+" or "-" key to set the flow rate.
• The flow rate ranges from 1 - 400 ml/h in 1ml increments.
• The flow rate slows down and stops shortly at 50 - 125 - 250 ml/h.
• Press the "DOSE=VOL" key and make sure the value is put at 0 ml indicated by the word
"CONT". This means the pump will run in a continuous mode.
• Start the pump by pressing "START/STOP".
• The arcs around the word "run" start rotating.
• The flow rate (ml/h) is displayed whilst functioning.
ZMNL900908_AW.indd 24
UK
PROGRAMMING THE PUMP
pump can be used for continuous or dose feeding.

CONTINUOUS FEEDING

+
400 ml/h
d
START/STOP
-
1 ml/h
• The arcs rotate when the pump is running.
• The remaining battery capacity (see paragraph: battery
operation).
• The wall plug symbol indicates that the charger is
plugged in.
-22-