Guidance and manufacturer's Declaration – Electromagnetic immunity
The Flocare
Infinity
®
below. The customer or the user of the Flocare
such an environment.
Immunity test
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
Recommended separation distances between portable and mobile RF communications equipment and the
The Flocare
®
Infinity
radiated RF disturbances are controlled. The customer or the user of the Flocare
tromagnetic interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the Flocare
power of the communications equipment.
Rated Maximum output
power of transmitter
(W)
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d
in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is
the maximum output power rating of the transmitter in Watts (W) according to the transmitter manufacturer.
NOTE 1
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorp-
tion and reflection from structures, objects and people.
ZMNL900908_AW.indd 33
enteral feeding pump is intended for use in the electromagnetic environment specified
Infinity
®
IEC 60601 test level
Compliance level
30 A/m
30 A/m
Flocare
Infinity
®
enteral feeding pump is intended for use in an electromagnetic environment in which
Infinity
®
Separation distance according to frequency of transmitter (m)
150 KHz to 80 MHz
d = 1.2
CD
P
0.12
0.38
1.2
3.8
12
enteral feeding pump should assure that it is used in
Electromagnetic environment – guidance
Power frequency magnetic fields should be
at levels characteristic of a typical location
in the typical commercial or hospital envi-
ronment.
enteral feeding pump
as recommended below, according to the maximum output
80 MHz to 800 MHz
d = 1.2
CD
P
0.12
0.38
1.2
3.8
12
-31-
®
Infinity
can help prevent elec-
800 MHz to 6 GHz
d = 2.3
CD
P
0.23
0.73
2.3
7.3
23
4.05.21 16:31