START/STOP
The "hold" mode is used to temporarily stop the flow of feed:
• to change the feeding program (installed parameters),
• to change the feed container,
• to administer medication without switching the pump off,
• to silence an alarm and correct problems.
After 3 minutes a two tone audible alarm sounds and the message "PAUSE" appears in the display.
Press "START/STOP" to stop the alarm and to extend the hold mode by a further 3 minutes or press
"START/STOP" twice (3 beeps will be heard) to resume programmed settings.
.
.
PURGER
Ensure the feeding set is not connected to the patients feeding tube when the pump is in the hold
mode: When the pump is in the hold mode: Press and hold the "PURGER" key for 2 seconds to
activate the "PURGER" maneuver. Release the button as soon as the pump generates a beep and
starts pumping at a flow rate of approx. 700ml/h. During this maneuver the message "PURGER"
-
appears in the display. The pump will automatically stop when the Flocare
p
set is completely filled with feed. However this "PURGER" maneuver can at anytime be stopped
by pushing the "PURGER" key a second time. The pump will return to the hold mode when the
"PURGER" maneuver is complete or stopped.
d
The pump has a liquid crystal display (L.C.D.) with large alphanumeric characters, smaller symbols,
words and a back light.
l
ZMNL900908_AW.indd 23
The feeding program (installed parameters) and total volume administered
since the last clearance will be retained in the pump's memory.
If the pump shuts down due to low battery voltage, the memory will be
retained for 24 hours.

"HOLD" MODE

To temporarily pause the pump, or switch into "hold" mode whilst operating,
press the "START/STOP" key once. Three beeps are heard and the run sym-
bol disappears.
The programmed flow rate (ml/h), volume (DOSE=VOL) and the administered
volume (ml) are retained.
FILLING THE PUMP SET
The "PURGER" function offered by the Infinity pump is used to fill a (new)
feeding set with nutrition (or water as the case may be). When the "PURGER"
function runs all alarms are deactivated helping avoid alarms being inadver-
tently triggered (e.g. air alarm).
The "PURGER" function should only be used to fill an empty (air-filled) feeding set. Use of the
"PURGER" function at any other time may cause the pump to incorrectly calibrate.