7. Configuration software WIKAsoft-TT
7.2 Configuration procedure
Steps 1 and 2 are carried out automatically when starting the software
1. "Loading instrument data"
2. "Loading configuration"
EN
3. [optional] Cancel write protection ("key" symbol at the bottom right)
4. Change the required parameters
→ Sensor/Measuring range/Error signalling etc.
5. "Save to the instrument"
6. [optional] Activate write protection
7. [optional] Print configuration history
8. [optional] Test: "Loading configuration" → Check configuration

7.3 Fault diagnosis

Here, in the event of an "error detected by the transmitter", the error
message is displayed.
Examples: Sensor break, permitted highest temperature exceeded, etc. In
normal operation, "No fault - No maintenance requirement" is displayed here.

7.4 Measured values

Line recorder - Here the measured value progression is represented in
the format of a chart recorder with a constant sampling rate in a defined
time interval (180 seconds) and a variable temperature axis.
The display purely serves as a functional check and for information.
An export of the data is not possible.
7.5 Configuring several instruments identically
First instrument
1. "Loading configuration"
2. [optional] Cancel write protection ("key" symbol at the bottom right)
3. Change the required parameters
4. "Save to the instrument"
5. [optional] Activate write protection
All following instruments
1. "Loading instrument data"
2. [optional] Cancel write protection
3. [optional] Change the required parameters, e.g. Tag number
4. "Save to the instrument"
5. [optional] Activate write protection
20
WIKA operating instructions model TR33