Interacoustics AD629 Instructions D'utilisation page 238

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Guidance and manufacturer's declaration - electromagnetic emissions
The AD629 is intended for use in the electromagnetic environment specified below. The customer or the
user of the AD629 should assure that it is used in such an environment.
Emissions Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
portable and mobile RF communications equipment and the AD629.
The AD629 is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the AD629 can help prevent electromagnetic interferences by
maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the AD629 as recommended below, according to the maximum output power of the
communications equipment.
Rated Maximum
output power of
transmitter
[W]
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where
P is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
Note 1 At 80 MHz and 800 MHZ, the higher frequency range applies.
Note 2 These guidelines may not apply to all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Compliance
Group 1
Class B
Complies
Class A Category
Complies
Recommended separation distances between
Separation distance according to frequency of transmitter
d = 1.17 √ ����
150 kHz to 80 MHz
0.12
0.37
1.17
3.70
11.70
Electromagnetic environment - guidance
The AD629 uses RF energy only for its internal
function.
Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.
The AD629 is suitable for use in all commercial,
industrial, business, and residential environments.
[m]
d = 1.17 √ ����
80 MHz to 800 MHz
0.12
0.37
1.17
3.70
11.70
d = 2.23 √ ����
800 MHz to 2.7 GHz
0.23
0.74
2.33
7.37
23.30