7 Care and maintenance
95°C
203°F
134°C
274°F
38
Operating Instructions BA ULTIMATE POWER + / BA ULTIMATE
The cleaning and disinfection equipment used must be approved by its
manufacturer for the cleaning and disinfection of dental instruments and comply
with EN ISO 15883-1 (e.g. 95°C (203°F) and 10 min. holding time).
For further details, refer to the operating instructions supplied with the unit.
1. Check whether the instrument is clean after processing.
2. If any contamination is present: repeat the procedure.
For further processing, the instrument should be dry and free
from any residues.
3. Blow the instrument out with max. 3 bar.
4. Manually maintain the mechanics [ → 31].
5. Manually maintain the pushbutton chuck [ → 33].
7.6
Sterilizing
1. Clean and disinfect the instrument [ → 35].
2. Apply spray to the instrument [ → 31].
3. Sterilize the instrument in the steam sterilizer with saturated water vapor.