Operation - B.A. International BA ULTIMATE POWER + Notice D'utilisation

Masquer les pouces Voir aussi pour BA ULTIMATE POWER +:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 82
6401553.book Seite 24 Montag, 10. September 2012 2:27 14

5 Operation

24
Operating Instructions BA ULTIMATE POWER + / BA ULTIMATE
5
Operation
Operation
NOTICE! Use only sharp and undamaged burs and diamond grinders. The burs
and diamond grinders must be clean to prevent dirt from being carried into the
clamping system.
CAUTION! A loose or partially removed bur instrument can detach itself from
the head or break off. This leads to a risk of injury! Therefore, only use the
instrument if the bur instrument is at least 10 mm deep and clamped securely in
place.
CAUTION! The handpiece must only be operated with the clamping system
closed.
CAUTION! Insufficient cooling leads to overheating of the cavity and damage
to the tooth substance. Ensure that the water content is > 50ml/min.
CAUTION! Do not pull the patient's cheek back with the contra-angle
handpiece while the motor is running. This would actuate the pushbutton, thus
creating a risk of burning the patient's oral mucosa.
CAUTION! Always operate the contra-angle handpieces with cooling spray
when marked with a red dot!
64 01 553 D3551.201.03.01.02

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Ba ultimateBa 200Ba 40Ba 60Ba 280Ba 250 ... Afficher tout

Table des Matières