Operating Instructions BA ULTIMATE POWER + / BA ULTIMATE
64 01 553 D3551.201.03.01.02
4.3
Before each patient
1. Purge the water paths and air channels for 20 - 30 seconds.
2.
CAUTION! Stop the motor.
Attach the instrument [ → 25].
3. Insert the bur instrument [ → 25].
4. Use a sufficient amount of cooling water (> 50 ml/min) [ → 29].
5. Operate only with filtered water (< 50 µm).
6. Check the nozzles for blockages and deposits of lime, for example, and clean
the nozzles if necessary.
CAUTION! Insufficient cooling leads to overheating of the cavity and damage
to the tooth substance. Ensure that the water content is > 50ml/min.
4 Preparation
23