Centrifugation Run With Program Linking; Deleting Program Linking; Cooling (Only In Centrifuges With Cooling); Standby-Cooling - Hettich ROTIXA 50 S Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

27.3

Centrifugation run with program linking

Press the
key twice in order to select the parameter PR-PART. The input field will be displayed inversely.
PROG
Set the program location for the first program (XX+) by means of the control knob
Press the
key. The centrifugation data for the selected program location will be displayed.
RCL
Press the
key. The rotation indicator
START
The run-up and braking steps for the program linking will be displayed:
First program (XX+)
x
x: Run-up step for the first program
PR-PART
Next program (+XX+)
x
x: Run-up step for the next program
PR-PART
Last program (+XX)
y
x
x: Run-up step for the last program
PR-PART
y: Braking step for the last program
When the time in the last program has expired, run-down will take place with the braking step of the last
program. If the centrifugation run is interrupted by pressing the
braking step of the program running at the time.
27.4

Deleting program linking

Press the
key in order to display the parameter PROG-Nr. The input field will be displayed inversely.
PROG
Set the program location for the first program (XX+) by means of the control knob
Press the
key. The centrifugation data for the selected program location will be displayed.
RCL
Press the
key twice in order to display the parameter PR-PART. The input field will be displayed inversely.
PROG
Press the
key twice.
STO
Press the
key.
PROG
28

Cooling (only in centrifuges with cooling)

The temperature set-point can be adjusted from -20°C to +40°C. On centrifuges with a heating/cooling option, the
temperature set-point is settable from -20°C to +60°C / +90°C. The lowest obtainable temperature is dependent on
the rotor (see Chapter "Anhang/Appendix, Rotoren und Zubehör/Rotors and accessories"). If the actual temperature
and the set temperature deviate more than 5 °C, the temperature value in the display will start to flash on and off.
28.1

Standby-cooling

With rotor standstill and closed lid the centrifugal chamber is cooled to the pre-selected temperature. The
temperature set-point is shown in the display.
28.2

Pre-cooling the rotor

For rapid pre-cooling of the unloaded rotor and the accessories, it is recommended that the centrifuge is run with the
continuous running settings and a speed of ca. 20% of the maximum rotor speed.
29

Heating (only on centrifuges with heating/cooling option)

During operation of the centrifuge the centrifuge chamber can be heated as required to the preselected temperature.
The heating is switched off when the rotor is stationary.
Risk of burning! The surface temperature of the heating element in the centrifugal chamber of the centrifuge
can reach 500°C / 932°F. Do not touch the heating element.
The plastic buckets, order no. 5091, may only be used for temperatures up to a maximum of 40°C.
will appear as long as the rotor is turning.
key, run-down will take place with the
STOP
EN
.
.
39/104

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Rotixa 50 rs

Table des Matières