X-ray Generator
High Voltage
Generator
(Assembly)
X-ray Tube
64
Item
Model
Rated output power
Duty Cycle
Cooling Protection
Inherent Filtration
Total Filtration
Type
Tube Voltage
Tube Current
Manufacturer
Model
Focal spot size
Anode heat contents
Maximum Anode Heat
Dissipation
Target Material
Target Angle
Inherent Filtration
X-ray Coverage
Tube Voltage
Tube Current
EzRay Air Portable (Model: VEX-P300) User Manual
Description
DG- S0101V1
Max. 0.2 kW
1:60 or more
(Exposure time: Interval time)
Thermistor ≥ 65 °C
1.8 mm Al / 65 kV
Min. 1.5 mm Al
Inverter Type
55-65 kV
1.0-3.0 mA
VATECH Co., Ltd.
V1-650304 (Stationary Anode type)
0.4 mm (IEC 60336)
Max. 2.7 kJ
200 W
Tungsten
12.5°
Min. 1.5 mm Al
70 mm at SID 200 mm
Max. 65 kV
Max. 3.0 mA