VATECH EzRay Air Portable Manuel D'utilisation page 57

Table des Matières

Publicité

5.3.8
Waiting Time Setting for the Power Down Mode
Perform this mode to set up the waiting time for the Power Down Mode. (default: 300
minutes)
Service Mode No.
n.21
To change the waiting time for the Power Down Mode,
Select n.21 by pressing the jog dial, then the default time "300" (300 minutes) is
1.
displayed on the Control Panel as shown in the following figure (right).
The waiting time can be modified from 1 to 999 minutes.
2.
Turn the jog dial to set the time, and save the setting.
3.
Check if it has been correctly saved.
EzRay Air Portable (Model: VEX-P300) User Manual