Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 16
► If damage is apparent or in case of doubt, do not continue using the
product.
► Take suitable measures if required (e. g. cleaning, repair, replacement,
inspection by the manufacturer or a specialist workshop etc.).
► Inform the patient.
Signs of changes in or loss of functionality during use
Reduced spring effect or decreased forefoot resistance are indications of
loss of functionality.
4 Scope of Delivery
The following single components and accessories are included in the scope
of delivery in the specified quantities and are available for reordering accord­
ing to their respective symbols as single components ( ), single compon­
ents with minimum order quantities ( ) or single-component packs ( ):
Fig.
Item no.
1
1
5 Preparation for Use
CAUTION
Incorrect alignment or assembly
Injuries due to fall resulting from damaged prosthesis components
► Observe the alignment and assembly instructions.
5.1 Selection of the Stiffness Version
The stiffness of the prosthetic foot is selected depending upon the body
weight and the preferred running style (short distance running or sprint/long
distance running). The recommended body weight ranges for each stiffness
version are listed according to running style in the following table.
Stiffness version
SPR-1
SPR-2
12 | Ottobock
1E90 Sprinter
Quant­
Designation
ity
   
1

Instructions for use

1
Prosthetic foot
Short-distance running
Body weight
[kg]
[lbs]
[kg]
[lbs]
40 to 52
90 to 115
53 to 63
116 to 140
Reference number
647G849
Long-distance running
Body weight
40 to 59
90 to 130
60 to 70
131 to 155