* Scaling
Auto or Manual
Scaling: Auto **
The spectrophotometer adjusts the scaling of the axes (minimum and maximum value of the axis) to the measured values
while measuring. The entire curve is always visible.
Scaling: Manual
Abs min.
Abs max.
The scaling of the axes (minimum and maximum value of the axis) is permanently set manually.
* Measurements/interval
1/interval or Max./interval
Here, it is possible to define how many measurements will be carried out per interval.
The setting affects the calculation of the slope of the individual intervals.
The displayed slope for an interval is determined as follows, depending on the profile:
Measurements/interval
1/interval
Max./interval
* Katalytic activity
YES or NO
Katalytic activity
NO
YES
* Factor
Factor multiplying the average slope report to the minute.
* Units
i.e.: µkat (10 characters maximum).
* Resolution
From 0 to 3
Number of digits, after comma, for the result.
1.45

METHOD SELECTION

- Select a method in the "Kinetcis" mode then press on <START·ENTER> key to confirm the method choice.
1.46

METHOD SAVING

Methods are automatically saved in memory.
Slopes
Difference of two measured values (2 points calculation)
Slope of the straight line determined in an interval by linear regression
Result
Total slope report to the minute
Total slope report to the minute multiplied by a factor