6.2 Error Codes List

The following list describes the various error codes which can appear on the display.
LoBat
Is indicated when batteries are weak. Approximately 20 minutes of operating time remain.
Error 1
Indicates an overload condition.
Error 2
Indicates an underload condition.
Error 7
EEPROM data incorrect.
Error 14
Zero exceeds ZERO% and cannot be zeroed.
Err 21
Calibration data does not match current full scale, Grad and Cal Point settings. Settings must
be restored or the Scale must be recalibrated using the current settings.

6.3 Service Information

If the Troubleshooting section does not resolve or describe your problem, you will need to contact an authorized Ohaus
Service Agent. For Service assistance in the United States, please call Aftermarket, Ohaus Corporation toll-free at (800)
526-0659. An Ohaus Product Service Specialist will be available to help you.

6.4 Accessories

Description
RS232 Interface Cable/SF42 Printer
RS232 Interface Cable/PC 25 Pin
RS232 Interface Cable/PC 9 Pin
Printer

6.5 Technical Data

Materials
ABS Housing
Keypad/display overlay polyester
Trooper Scale
Ohaus Part No.
80500573
80500431
80500433
SF42