Information Codes; Error Codes - ENBIO S Mode D'emploi

Masquer les pouces Voir aussi pour S:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 62

9.2 Information codes

9.3 Error codes

The following table includes the error codes that may be displayed during
the use of the ENBIO steriliser.
Error code
Description
1
"Chamber over
temperature"
2
"Steam gen. over
temperature"
3
"Process over tem-
perature"
4
"Overpressure error"
Screen for overpressure or under-
pressure
resulting
from
chamber cooling processes.
Message resulting from interruption
of the process after the sterilization
stage - during drying.
Recommendations
Maximum tempera-
Contact the service
ture in the chamber
exceeded
Excessive steam
Contact the service
generator
temperature
Excessive process
Contact the service
temperature
Pressure error
Contact the service
5
"Sterilization pressure
too low"
6
"Sterilization temp.
natural
too low"
7
"Too high pressure
during drying"
8
"Too many steam
pulses/no water"
9
"Drainage error"
10
"Chamber heating
error"
11
"Steam generator
heating error"
12
"Prevacuum fail/check
condensate outlet"
13
"Power failure"
14
"Pressure during
standby"
15
"Locking door error"
16
"Unlocking door error" Door unlocking error
17
„Valve V3 / HEPA filter
error"
18
"Pressure sensor error" Pressure sensor error
19
"USB disc error /
Change disc"
31
"Internal flash error"
Pressure too low
Check water level and
during sterilisation
connection.
Contact the service
Sterilization
Check water level and
temperature too low
connection.
Contact the service
Pressure too high
Check if the outlet
during drying
tube is not submerged
in water.
Contact the service
Too many steam
Check water supply
impulses.
level and tubing
No water supply.
connections.
Contact the service
Drainage clogged
Check wastewater
level and tubing
connections.
Contact the service
Chamber heating error Contact the service
Steam generator error Contact the service
Vacuum pump/
Check wastewater
drainage error
level and tubing
connections.
Contact the service
Temporary power loss
Confirm error.
during operation
Pressure exceeded
Confirm error.
during standby
Contact the service
Door lock error
Contact the service
Contact the service
V3 valve / HEPA filter
Check filter cleanli-
error
ness/replace filter.
Contact the service
Contact the service
Write error on the
Write error on the
pendrive - media
pendrive - damage to
damage
the media. Rip content
from the current
pendriva - purchase of
a new one and use of
a new one
Memory error
Contact the service
56

Hide quick links:

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Pro

Table des Matières