System information

Comparing meter results to
laboratory results
Results obtained from the OneTouch Verio Reflect
Meter and laboratory tests are reported in plasma-
equivalent units. However, your meter result may differ
from your lab result due to normal variation. A result
from your OneTouch Verio Reflect
Meter is considered
accurate when it is within 0.83 mmol/L of a laboratory
method when the glucose concentration is lower than
5.55 mmol/L and within 15% of a laboratory method
when the glucose concentration is 5.55 mmol/L
or higher.