WIKA T32. S Serie Guide De Sécurité page 7

Masquer les pouces Voir aussi pour T32. S Serie:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

2. Safety
The temperature transmitter achieves, for all recognised connections for temperature
sensors, an SFF (Safe Failure Fraction) sufficient for SIL 2 of > 90 %.
The instrument generates a current signal in the approved measuring mode of a
nominal 4 ... 20 mA, that is dependent upon the sensor signal. The effective range of
the output signal limited to a minimum of 3.8 mA and a maximum of 20.5 mA (factory
setting in basic configuration).
WARNING!
Do not exceed the specifications for model T32.xS given in the data
sheet and operating instructions. In order to ensure a safe functionality
of the current output, the correct terminal voltage must be present in the
instrument.
The following terminal voltage limits apply:
Instrument model
T32.1S.000-S
T32.3S.000-S
T32.1S.0IS-S
T32.3S.0IS-S
T32.1S.0NI-S
T32.3S.0NI-S
T32.1S.0IC-S
T32.3S.0IC-S
WARNING!
Only the temperature sensors listed in chapter 2.1 are approved for use in
a safety-related application.
The following sensors and operating modes are NOT allowed for
operation in a safety-relevant application:
Potentiometers
Other resistance sensors
Other mV sensors
Differential mode in duplex sensor operation
WIKA safety manual temperature transmitter T32.xS
Terminal voltage limits
DC 10.5 ... 42 V
DC 10.5 ... 30 V
DC 10.5 ... 40 V
DC 10.5 ... 30 V
GB
7

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

T32.1sT32.3s

Table des Matières