TECNO-GAZ
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Operation and
maintenance manual
IEC 60601 test
level
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2,5GHz
MU_025
Rev.2
Compliance
Electromagnetic environment-
level
Portable and mobile RF
communications equipment should
be used no closer to any part of the
Product, included cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation
distance
d = 1,2√P 150 KHz to 80 MHz
d = 1,2√P 80 MHz to 800 MHz
d = 2,3√P 80 MHz to 2,5 GHz
3 Vrms
where P is the maximum output
power rating of the transmitter in
3 V/m
watts (W) according to the
transmitter manufacture and d is
the recommended separation
distance in meters (m).
Field strengths from fixed
transmitters, as determined by an
electromagnetic site survey, should
be less than the compliance leave
in each frequency range.
Interference may occur in the
vicinity of equipment marked with
the following symbol.
04/02/16
EN
Page 25 of 28