Product Liability And Warranty; General Information; Scope Of Delivery; Intended Use - KLS Martin group NON-STICK red Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Instructions for Use
Bipolar Forceps NON-STICK
1

Product Liability and Warranty

1.1

General Information

We thank you for having decided to buy a KLS Martin product. This product carries the
CE-mark, which means that it satisfies the essential requirements laid down in the EC Directive
concerning medical devices.
We are the manufacturer of this product:
Gebrüder Martin GmbH & Co. KG
A company of the KLS Martin Group
KLS Martin Platz 1 · D-78532 Tuttlingen · Germany
Postfach 60 · D-78501 Tuttlingen · Germany
Tel. +49 7461 706-0 · Fax +49 7461 706-193
info@klsmartin.com · www.klsmartin.com
1.2

Scope of Delivery

Bipolar forceps NON-STICK
Item no. 80-982-12-04 to 80-999-24-04
1.3

Intended Use

The KLS Martin bipolar forceps NON-STICK
grasping and manipulating and tissue. A suitable bipolar cable must be used to connect these
products to the appropriate output of an electrosurgical unit that uses bipolar coagulation
current.
The maximum output voltage of the generator may not exceed 500 Vp. The appropriate output
power setting depends on the specific application, the type of forceps used and the selected
coagulation current. In addition, however, the relevant instructions provided in the user man-
ual of the electrosurgical unit must be observed and followed as well.
Suitable bipolar cables:
Products with item numbers 80-287-33-04 / 53-04, 80-286-30-04 / 50-04,
80-287-89-04 / 90-04, 80-287-70-04, 80-291-40-04, 80-286-40-04, 80-293-40-04.
WARNING
Danger of injuries due to insufficient qualification of the operator!
Safe usage of the bipolar forceps NON-STICK
the methods and application of electrosurgery.
Electrosurgical instruments may only be used by qualified individuals specially trained or
instructed in their use!
Revision 3
r e d
r e d
r e d
r e d
are intended for coagulating, dissecting,
requires that the operator is familiar with
19

Publicité

Table des Matières
loading

Table des Matières