Recommended Separation Distances; Cable Lengths - VOCO Celalux 2 Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 14
Note 2: These specifications may not be applicable in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and persons.
(a): The electromagnetic field strengths of fixed radiofrequency emitters, such as base stations for mobile telephones
(cellular/cordless), mobile radios, amateur radios, AM/FM radio broadcasts and TV broadcasts cannot be
determined exactly by theory. To assess the electromagnetic environment due to fixed radiofrequency emitters,
an electromagnetic environment measurement must be made. If the measured radiofrequency field strength in
the immediate environment where the product is used exceeds the compliance level specified above, the
performance of the product must be tested to verify whether it conforms to the specifications. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocating the product.
(b): In the 150 kHz to 80 MHz frequency range, the electromagnetic field strengths must be less than 3 V/m.

10.4 - Recommended separation distances

The C
2 is intended for use in an electromagnetic environment in which radiated radiofrequency disturbances
ELALUX
are controlled. The C
ELALUX
minimum distance between portable and mobile radiofrequency communications equipment (emitters) and the
C
2, according to the maximum output power of the equipment, as recommended in the table below.
ELALUX
Rated max. power
of the emitter (W)
0.01
0.1
1
10
100
For emitters rated at max. power not listed above, the recommended separation distance d in meters (m) can be estimated using
the equation applicable to the frequency of the emitter, where P is the max. power rating of the emitter in watts (W) according
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These specifications may not be applicable in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and persons.

10.5 - Cable lengths

Cables and accessories
Maximum length
charger power cord
2 user and/or installer can help prevent electromagnetic interference by maintaining a
Separation distance in meters (m) according to emitter frequency
150 KHz to 80 MHz
d = 1.2
0.12 m
0.38 m
1.2 m
3.8 m
12 m
RF emission, CISPR 1 - Class B
Immunity to magnetic fields: IEC61000-4-8.
Immunity to electrostatic discharge: IEC61000-4-2
Immunity to electrical fast transients/bursts: IEC61000-4-4
<3m
Immunity to surges: IEC61000-4-5
Immunity to voltage dips, short interruptions and voltage variations:
IEC61000-4-11
Immunity to conducted disturbances induced by radiofrequency fields: IEC61000-4-6
Immunity to radiated radiofrequency electromagnetic fields: IEC61000-4-3
80 MHz to 800 MHz
d = 1.2
0.12 m
0.38 m
1.2 m
3.8 m
12 m
the manufacturer.
Complies with:
10
800 MHz to 2.5 GHz
d = 2.3
0.23 m
0.73 m
2.3 m
7.3 m
23 m

Publicité

Chapitres

Table des Matières
loading

Produits Connexes pour VOCO Celalux 2

Table des Matières