Cleaning
Protective glass
Clean the protective glass with a microfiber cloth to
make sure all light reaches the object.
Note! Solvents are not allowed.
Filter
Clean the filter (no. 5 in chapter 3.1) regularly. Frequency
is depending on the curing device environment. If
the filter is not clean, the IRT UV SmartCure will be
overheated and stops.
Remove filter as shown below.
Clean by gently blowing with compressed air or by
rinsing in water. Note! Solvents are not allowed.
715098 rev 3
715128
8. Trouble shooting
High temperature / Overheating
1. Wait until temperature drops.
2. Check ambient temperature. Maximum is 40°C.
3. Check filter. Clean filter as described in chapter 7.3.
4. Clean protective glass as described in chapter 7.3.
Service symbol with code
715128
1. Unit is broken.
2. Check product identity label, see chapter 6, and refer
3. Contact Hedson Technologies Support for service:
Battery uncharged
Change battery as shown in chapter 7.2.
EN
to Serial No and Produced when contacting Hedson
support, see below.