Validation; Maintenance And Cleaning; Temperature/Pressure Chart - VWR VAPOUR-Line eco 25 Manuel D'instructions

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 41
sterilisation time a buzzer is activated
5. In the cooling process, when the vessel temperature goes below 100 °C the LED END CYCLE will light up
6. Once the cooling process is completed, open the lid with the handle
7. Switch off the autoclave via the main switch (Pos. 1 page 7)
NOTE:
Press the STOP button to end a running cycle.
WARNING! STOP button is an emergency button. Pushing it will abort the cycle and will open the
valves. We don't recommend you push it unless a problem occurs during the cycle.
The same thing will happen if the instrument is turned off by the main switch.
The set values are stored in the memory and will not be lost after switching the instrument off.

Validation

The autoclave should be validated weekly with chemical indicator, and monthly with biological indicators.

Maintenance and cleaning

The autoclave should be inspected annually according to "Good Laboratory Practice".
Daily cleaning of the autoclave will prevent most future faults.
Clean the autoclave with a mild detergent using a wet soft cloth. Replace the vessel water daily with new, fresh
water. Long-term use of dirty water can cause premature failure of the heating element.
WARNING
Do not use any cleaning product containing halogens or other aggressive substances to clean the
autoclave as this can cause rust.

Temperature/pressure chart

Temperature (°C)
100
Pressure (bar)
0
102
105
109
112
115
0,1
0,2
0,4
0,5
0,7
119
120
128
134
0,9
1,0
1,5
2,0
14

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Vapour-line eco 25 p481-0671481-0708

Table des Matières