Xylem PolluStat Instructions De Montage Et Manuel D'utilisation page 95

Compteur d'énergie thermique à ultrasons
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

11 Information Messages
When the instrument has detected an information message, the message symbol is displayed:
The specific message can be found at menu item 6 'Information message' in level 1 / main loop (see section 8, Display). The
message code is displayed alternately in binary and hexadecimal form.
The instrument recognizes eight message causes, which can also occur in combination with each other.
Hexadecimal display
H 80
H 40
H 20
H 10
H 08
H 04
H 02
H 01
„Temperature sensor 1" is the right temperature sensor (viewed from the front).
Example: Temperature sensor 1 cable break
Message
Bit
Display location
Alternating binary
message displayed
(LCD)
When a message
appears in the standard display (total heat energy), with the exception of the messages
-
Low battery (H 80)
-
Reset (H 40)
-
Error in flow measurement system (H 10; in the case of air in the ultrasonic measuring tube), the instrument
must be exchanged and sent to the supplier for examination.
Page 95 of 108
Article no.: 1080619046
Description
Low battery
Instrument has been reset
Electronics defective
Error in flow measurement system
Temperature sensor 2 short circuit
Temperature sensor 2 cable break
Temperature sensor 1 short circuit
Temperature sensor 1 cable break
7
6
5
1
2
3
4
3
2
4
5
6
2020_07_27
Binary display
1 at first place
1 at second place
1 at third place
1 at fourth place
1 at fifth place
1 at sixth place
1 at seventh place
1 at eighth place
1
0
Alternating hexadecimal
7
8
message displayed (LCD)
Subject to technical change! Errors excepted.

Publicité

Table des Matières
loading

Table des Matières