Degree of protection against harmful ingress of liquids: Ordinary equipment
without protection against ingress of water.
Electro-Magnetic Compatibility: Group I, Class B
6 Over-limit Indication
6.1 Limit settings
- SpO
low limit setting range: 85% ~ 95%.
2
- Pulse Rate limits setting range:
High: 100bpm--240bpm Low: 30bpm--60bpm
During the measurement, if the measured value exceeds the preset value,
the alert beeping sound will be activated, the value that is over-limit will blink
at the same time.
6.2 Alert sound mute setting
- During the measurement, if the alert sound is set to on, short time press "
" recall/mute key, then the alert sound will mute for 90 seconds, but the
over-limited value still keeps blinking. At this moment, the sound indication
icon " " becomes "
alert sound will be activated again.
- During the measurement, if the probe is off or disconnected, the message
"Check Probe" shows on the display screen. The alert sound starts (interval is
5 seconds) and lasts for about 1 minute. If the probe is still off, the Oximeter
will power off automatically.
7 Accessories
1. A probe
2. A holder
3. Battery "AA" × 3
4. User Manual
5. Quality Inspection Certificate
6. A data cable (optional)
7. Oximeter Data Manager software (optional)
Note: The accessories are subject to change. See the Packing
List for detailed items and quantity.
ENGLISH
". If this alert event persists over 90 seconds, then the
34