Les langues disponibles

Les langues disponibles

4. To save the settings to the internal memory of the instrument, select Save Settings to Permanent
Flash.
Note: If this option is not selected, the instrument still operates with the new settings. If the power to the
instrument is lost, the new settings are lost and the instrument goes back to the previous settings.
5. Click OK.
Configure the raw data text output mode (PM4000 mode)
1. From the main menu, select the Edit Settings icon.
2. Select the Counter tab.
3. Select Raw Data Output Text Mode to enable the feature.
4. Click OK to save the selection.
Configure the alarms
1. From the main menu, select the Edit Settings icon.
2. Select the Alarms tab.
3. Make changes to these fields:
• Standard (ISO, NAS or SAE)
• Alarm Direction (Clean to Dirty or Dirty to Clean)
• Code Limits (4, 6, 14 or 21 µm) If NAS is selected as the Standard, only the code limits for
6–14 µm and 14–21 µm will be available.
4. To open the contact when an alarm occurs, select Open Contact on Alarm.
5. To close the contact when an alarm occurs, do not select Open Contact on Alarm.
6. To save the settings to the instrument internal memory, select Save Settings to Permanent Flash.
Note: If this option is not selected, the instrument still operates with the new settings. If the power to the
instrument is lost, the new settings are lost and the instrument goes back to the previous settings.
7. Click OK.
Configure the general settings
1. From the main menu, select the Edit Settings icon.
2. Select the General tab.
3. Make changes to these fields.
• Display ISO Codes (All Codes or Highest Code)
• Display NAS Codes (All Codes or Highest Code)
• Display SAE Codes (All Codes or Highest Code)
4. To calculate average counts, select Calculate Average Counts and adjust the number of samples.
5. To change the output folder, click the button next to the Log File Output Folder field and select
the desired folder. As an option in Windows
text box.
6. Click OK.
Configure the baud rate
1. From the main menu, select the Edit Settings icon.
2. Select the Counter tab.
3. Click on the drop-down arrow in the baud rate box and select the baud rate.
4. Click OK to save the selection. The new setting will be set in the instrument and written to the
(permanent flash) internal memory. The new setting starts when the instrument is power-cycled.
14 English
®
®
Explorer
, move a copy of the output folder into the