Function; 3-Port Isolation Of Inputs, Outputs And Power Supply - Knick PolyTrans P 32000 P0 Série Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

The universal transmitter periodically samples signals from thermo-
couples, resistance transducers, strain gauges, resistive sensors, etc.
These signals are converted into output signals proportional to the
measured values. The characteristic can be controlled via the IrDA
interface. The output signal can be a voltage or a current.
3-port isolation with protective separation up to 300 V AC/DC accord-
ing to EN 61140 ensures optimum protection of personnel and equip-
ment as well as unaltered transmission of measuring signals.

3-Port Isolation of Inputs, Outputs and Power Supply

Inputs
Power supply
Basic Insulation
Working voltage
Overvoltage category
Pollution degree
Protective Separation According to EN 61140 by
Reinforced Insulation According to EN 61010-1
Working voltage
Overvoltage category
Pollution degree
Outputs
up to 300 V AC/DC
II
2
up to 300 V AC/DC
II
2

Function

WARNING!
For applications with
high working voltages,
take measures to prevent
accidental contact and
make sure that there
is suffi cient distance
or insulation between
adjacent devices.
43

Publicité

Table des Matières
loading

Table des Matières