FILLING THE FILLING TANK
- Only use distilled water (see technical specifications on page 62).
- When the LCD display shows the message "Proceed to fill H2O" fill the filling tank with
distilled water, choosing from one of two of the following modes:
AUToMATIC FILLING
- Insert the tube part, with the quick coupling, in the specific water filling joint (12 Fig. 2)
and insert the other end in the canister.
- Press the (+) button to start the water filling pump.
- Press the (-) key (\\) to stop the operation.
- Filling is automatically blocked once the maximum level of the filling tank is reached.
- Automatic filling of feeding tank from water mains.
MANUAL FILLING
- Insert the tube part with the quick coupling in the hole (14 Fig. 3).
- Insert the funnel into the other end of the tube and pour max 2 litres of water.
EMPTYING THE DISCHArGE TANK
- Insert the side of the tube without the quick coupling in a container to collect the discharged
water.
- Insert the quick coupling in the hole 11 Fig. 2 and position the container under the autoclave
to empty the tank.
- Water must be discharged in compliance with the standards in force on waste
disposal.
- Automatic emptying of discharge tank (16 fig. 3)
PLACING MATErIAL To STErILISE IN THE CHAMBEr
Wrap the load very carefully determining the compatibility of the medical devices to wrap
with the pre-chosen wrapping.
The material used for packaging must ensure:
- Compatibility with the sterilisation process.
- Compatibility with the labelling system.
Always insert a chemical indicator with the load to verify the positive outcome of the
sterilisation thermodynamic cycle.
Arrange the well-cleaned material on the trays, distributing it uniformly.
The chamber can have an inserted maximum of:
wrapped material
unwrapped material
porous load
The wrapped objects must be placed on trays with the transparent part
of the wrapping turned downwards.
"ProGrAMMED STArT" FUNCTIoN
From the menu the user can access the sub-menus "Instant start - Programmed Start"
a- Select Instant start to access the sterilization cycle or the test cycle.
b- Select Programmed Start to access the menu for setting the start time. The user can
set the hours and the minutes by entering every digit using the keys +/- selecting with
the Enter key
Once you have set the Start time, access the menu for selecting the sterilisation cycle or
the test cycle.
Model 16 L
2,5 kg
4 kg
1,3 kg
. To confirm the setting go to Save and confirm with the Reset key.
Model 21 L
3,5 kg
6 kg
1,6 kg
45