93

Initial Filling or Rinsing of the Entire Interchangeable Unit

2.6.1
While the initial filling or rinsing programme is being run, please place a sufficiently dimensioned waste
vessel under the titration tip.
Initial filling of the interchangeable unit is done using the «rinsing» program.
Fig. 10
On the main menu (Fig. 10), press <MODE> to navigate to the methods/system.
Pressing <↑> twice will take you to the «Rinsing» selection immediately (Fig. 11).
Fig. 11
Confirm the selection by pressing <ENTER>/<OK>.
At this point you can select the number of rinsing cycles (Fig. 12).
Initial filling requires a minimum of two rinsing cycles!
Fig. 12
Confirm the selection by pressing <ENTER>/<OK>.
At this point you can select the number of rinsing cycles (Fig. 13 - Fig. 16).