Xylem Si Analytics TitroLine 5000 Mode D'emploi page 152

Masquer les pouces Voir aussi pour Si Analytics TitroLine 5000:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

152
Command
Description
aaAA
automatic allocation of device address
aaMC1...XX
choosing a method
"filling burette". Aufsatz wird gefüllt.
aaBF
aaBV
output of dosed volume in ml
aaDA
dose volume without filling, with adding the volume
aaDB
dose volume without filling, reset of the volume
aaDO
dose volume with filling, without adding the volume
aaGF
filling time in seconds (min is 20, default 30)
"exit" function.back to main menu
aaEX
aaFP
pH measurement function
aaFT
temperature measurement function
aaFV
mV measurement function
dosing speed in ml/min (0.01 – 100 ml/min)
aaGDM
filling time in sec (adjustable 20 – 999 seconds)
aaGF
aaGS
output serial no. Of device
aaLC
output of the CAL parameters
aaLD
output of the measurement data
aaLR
output report (short report)
aaM
output of the preset measurement value (pH/mV/ug)
aaLI
output method content
aaRH
request of identification
aaRC
send last command
aaRS
report status
possible answers are:
"STATUS:READY" for ready
"STATUS:dosing" dosing
"STATUS:filling" filling
"ERROR:busy" if no interchangeable unit has been attached
aaSM
start selected method
aaSEEPROM
EEPROM reset to factory defaults
aaSR
stop the actual function
aaSS
titration start with the transfer of the pH end value
aaVE
software
Reply
aaY
aaY
aaY
aa0.200
aaY
aaY
aaY
aaY
aay
aay
aay
aay
aaY
aaY
aaGS08154711
aaY
aaY
aaM7.000
®
aaIdent: TitroLine
5000
aa"last command"
aaStatus:"text
aaY
aaY
aaY
aaY
Version number of the
aaVersion

Publicité

Table des Matières
loading

Table des Matières