Specifications
150mA 40ms Trip Test
This is a stand alone test at 150mA for 40ms
Displayed Range
0,1ms to 40ms
There is no associated
1
2
Power Supply
6 x 1,5V Alkaline cells type LR6 or 1,5V nickel cadmium,
(NiCd) or nickel metal hydride, (NiMH)
Fuses
Non replaceable
2 x 7A (SIBA 70-065-63)
The 7A fuses protect the instrument and are not
replaceable by the user.
Safety
Meets the requirements for double insulation to
IEC61010-1 (1995), EN61010-1 (1995) Installation
Category III***, up to 300V to earth and 400V phase to
phase, without the need for separately fused test leads.
If required, fused test leads are available as an optional
accessory.
Complies with the relevant parts of EN 61557:1997-02
as detailed below.
*** Relates to the transient overvoltages likely to be met
in fixed wiring installations. E.M.C. Meets EN61326-1
(1997)
Environmental Protection
IP54
The instrument is designed for
indoor and outdoor use.
∆n test or Delay.
I
Temperature Range:
Operating
-5°C to +40°C up to 90% RH
Storage
-25°C to +65°C up to 90% RH
General
Dimensions
230mm x 114mm x 62mm
Weight
920g with battery but without
leads and test & carry case
Cleaning
Wipe the disconnected
instrument with a clean cloth
dampened with soapy water or
Isopropyl Alcohol (IPA).
IEC 61557/EN 61557
Complies with the following parts of EN 61557,
Electrical safety in low voltage systems up to 1000V
a.c. and 1500 V d.c. – Equipment for testing, measuring
or monitoring of protective measures:-
Part 1 - General requirements
Part 3 - Loop resistance
Part 6 - Residual current devices (RCDs)
37