Problem
The control
unit cannot be
switched on.
The cuffs open
by themselves
during use.
Possible cause
The timer function is
activated but no period
of operation is set.
The control unit is
not connected to the
mains.
There is no mains
voltage.
The fuse in the control
unit has been tripped
due to overvoltage
(e. g. lightning strike).
In the event of over-
load, the zips on the
cuffs open so as to
prevent a blockade of
the arteries.
Fix
Check the position of the main
switch. If it is in position II, the
timer function is activated. In
order to start the control unit you
have to set the rotary knob of the
timer to the required period of
operation.
If you do not wish to use the timer
function, press the main switch to
position I.
Insert the mains plug into the
socket.
Try another mains socket.
Check the fuse of your mains
connection.
Contact the customer service
department.
Use cuff extensions. These are
available at
www.venenwalker.com.