Section 3 Installation

Temperature Sensor

Tube and Coil

CAUTION
Mounting
Accessory
(EK45 only)
CAUTION
3-2

Fuses

All units are equipped with automatic thermally-triggered fuses. A red
marking indicates a triggered fuse. Allow the unit to cool, approximately five
minutes, before pressing the fuse reset on the rear of the unit.
If the fuse triggers again the cooler has a defect. Contact our Customer
Service and Support.
The Pt100 sensor should be located in the bath in the
direct vicinity of the evaporator coil.
A temperature accuracy of approximately 1 to 2°C is
possible. Any temperature within the working temperature
range can be set. A heater is necessary for applications
which require temperatures above the respective ambient
temperature.
Never sharply bend the evaporator tube and coil or form it into a small
radius.
Damage caused by an excessive bending of the cooling coil is not covered by
our warranty.
Evaporator hose − the smallest permissible bending radius is 125 mm.
Evaporator coil (cooling coil) − the smallest permissible bending radius for
the EK90 is 40 mm (one bend), or 110 mm (more than one bend).
To fit the evaporator in the optional clamping bracket
slide the metal plates supplied with the bracket into
the slot above the evaporator coil.
Insert the cooling element through the rectangular
opening
Attach the inserted plates using four screws.
Do not kink the evaporator hose.
Ω
3
2
1
4
Thermo Scientific