Dentsply Sirona Calamus Dual A1300 Protocole D'utilisation page 152

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 83
DISINFECTION, CLEANING AND STERILISATION
Operation
9
Storage
B. Filling material
Operation
1
Disinfection
32/40
Operating mode
• Keep devices
in sterilization
packaging in a
dry and clean
environment
Operating mode
Immerse the
obturation devices
in NaOCl (2,5%
at least) during
5 mn at ambient
temperature.
BENCALDDFUWEB / Rev.04 / 02-2019 (Old ZF 1902103.EN)
A3. Contra Angle
A2. Implantable devices
Following uses
A1. Instruments
First use
Warning
• Sterility cannot be guaranteed if
packaging is open, damaged or
wet.
• Check the packaging and the
medical devices before using
them (packaging integrity, no
humidity and validity period).
Do not use disinfecting solutions
containing Phenol or any products which
are not compatible with the treated filling
material
(See "General
X X X X
Warning
recommendation").

Publicité

Table des Matières
loading

Table des Matières