DISINFECTION, CLEANING AND STERILISATION
Operation
4a Automated
Cleaning
with washer-
disinfector
OR
4b Manual
Cleaning or
assisted by
an ultrasonic
device
BENCALDDFUWEB / Rev.04 / 02-2019 (Old ZF 1902103.EN)
Operating mode
• Place the devices
in a kit, support
or container
to avoid any
contact between
instruments or
posts.
• Put them in the
washer-disinfector
(Ao value > 3000
or, at least 5 min
at 90°C).
• Place the devices
in a kit, support
or container
to avoid any
contact between
instruments.
• Immerse in the
disinfecting
solution with
cleaning
properties,
assisted by an
ultrasonic device if
suitable
A3. Contra Angle
A2. Implantable devices
Following uses
A1. Instruments
First use
Warning
• Discard any instruments with
large obvious defects (broken,
bent).
• Avoid any contact between
instruments or posts when
placing in the washer-disinfector
use kits, supports or container.
• Follow instructions and observe
concentrations given by the
manufacturer
(See "General
recommendation").
• Use only approved
washer-disinfector according to
EN ISO 15883, maintain and
calibrate it regularly.
• No visible impurities should be
observed on the instruments.
• Discard any instruments with
large obvious defects (broken,
bent, and twisted).
• Follow instructions and observe
concentrations and time given by
the manufacturer
(See "General
recommendation").
• The disinfecting solution should
be aldehyde free and without di-
or triethanolamines as corrosion
inhibitor.
X X X X
X X X
29/40