DISINFECTION, CLEANING AND STERILISATION
4. Limitations and restrictions on reprocessing:
The individual DFU indicates if the useful life of a device might be
reduced by the number of reprocessing cycles.
Furthermore, the appearance of defects such as cracks, deformations
(bent, twisted), corrosion, loss of colour coding or marking, are
indications that the devices are not able to fulfil the intended use with
the required safety level.
5. Single use marked instruments are not approved for re-use.
6. The water quality has to be convenient to the local regulations
especially for the last rinsing step or with a washer-disinfector.
7. Tungsten carbide burs, plastic supports, hand instruments and NiTi
instruments are degraded by Hydrogen Peroxide (H
8. NiTi Instruments are degraded if immerged more than 5 minutes in a
solution of NaOCl at more than 5%.
9. Instruments made of aluminium are degraded in presence of caustic
soda solutions with mercury salt. Do not use acid (pH < 6) or alkaline
(pH > 8) solutions.
10. The washer-disinfector is not recommended for instruments made of
aluminium, tungsten carbide or carbon steel.