DISINFECTION, CLEANING AND STERILISATION
Operation
5
Rinsing
6
Inspection
30/40
Operating mode
• Abundant rinsing
(at least 1 min)
• Inspect devices
and sort out those
with defects.
• Assemble the
devices (stops)
BENCALDDFUWEB / Rev.04 / 02-2019 (Old ZF 1902103.EN)
A3. Contra Angle
A2. Implantable devices
Following uses
A1. Instruments
First use
Warning
• Use quality water in accordance
with local regulations.
• If a disinfecting solution
contains a corrosion inhibitor,
it is recommended to rinse the
instruments just before the
autoclaving.
• Dry on a single use non-weaved
cloth, or with a drying machine or
filtered compressed air.
• Dirty instruments must be
cleaned and disinfected again.
• Discard instruments which show
any deformations (bent, twisted),
damages (broken, corroded) or
defects (loss of colour coding or
marking) affecting the resistance,
the safety or the performance of
the instrument or posts.
• Protect carbon steel bur with
corrosion inhibitor before
packaging.
• For Contra Angle: lubricate the
device with an adequate spray
before packaging
X X X
X X X X