Dentsply Sirona Calamus Dual A1300 Protocole D'utilisation page 151

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 83
DISINFECTION, CLEANING AND STERILISATION
Operation
7
Packaging
8
Sterilization
BENCALDDFUWEB / Rev.04 / 02-2019 (Old ZF 1902103.EN)
Operating mode
• Place the devices
in a kit, support
or container
to avoid any
contact between
instruments or
posts and pack
the devices in
"Sterilisation
pouches".
• Steam sterilisation
at:134°C / 273°F
during 18 min.
A3. Contra Angle
A2. Implantable devices
Following uses
A1. Instruments
First use
Warning
• Avoid any contact between
instruments or posts during
sterilization. Use kits, supports or
containers.
• Check the validity period of the
pouch given by the manufacturer
to determine the shelf life.
• Use packaging which are
resistant up to a temperature of
141°C (286°F) and in accordance
with EN ISO 11607.
• The instruments, posts and
the plastic supports must be
sterilized according to the
packaging labelling.
• Use only autoclaves that are
matching the requirements of
EN 13060, EN 285.
• Use a validated sterilisation
procedure according ISO 17665
• Respect the maintenance
procedure of the autoclave
device given by the
manufacturer.
• Use only this recommended
sterilization procedure.
• Control the efficiency (packaging
integrity, no humidity, colour
change of sterilisation indicators,
physico-chemical integrators,
digital records of cycles
parameters).
• Traceability of procedure records
X X X X
X X X X
31/40

Publicité

Table des Matières
loading

Table des Matières