Testing the NightWatch

The base station will emit 1 short beep every two seconds as soon as you remove the activated
arm module from the charger. This will stop as soon as the NightWatch can monitor a reliable heart
rate. This confi rms that the NightWatch is working correctly and that the volume has been set as
required.
If you want to check the volume of the alert that the base station would emit in the event of an
epileptic seizure, press either of the two volume buttons.
6.5

Connection to an external call system (optional)

EN
A
The NightWatch can be linked to an external call system. This enables the NightWatch alerts to be
transmitted to special systems intended to alert caregivers who are located too far away from the
base station. This can be done via the RJ-11 (relay) or a RJ-45 (API) connection to the base station.
If you would like to create such a link, please contact your supplier so that they can confi gure your
NightWatch appropriately and help you create the correct link.
PLEASE NOTE: Be sure to regularly check the link with an external call system by triggering an alert by
the NightWatch and subsequently verifying that the alert is correctly received and transmitted to and/
or responded to by the call system.
You can test whether an alert is transmitted correctly by stimulating an epileptic seizure alarm. To
do this:
» Disconnect the arm module from the charger.
» The base station will emit an alert that a reliable heart rate has not yet been detected. However,
» Hold the arm module sensor still against your skin at a spot where the heart rate can be
» Wait until the base stations stops beeping and LED 5 is blinking green. This indicates that the
80
NightWatch user manual v4.0 v3.indd 80
NightWatch user manual v4.0 v3.indd 80
A
B
B
= RJ-45 port
= RJ-11 port
this start alert will not be transmitted via the link as a technical alert.
monitored easily and clearly. Such as your cheek, for example.
arm module has been able to monitor a reliable heart rate.
C
C
= Power supply
v4.0
NightWatch
15-07-20 16:12
15-07-20 16:12