Surface Disinfection; Removal Of Radioactive Contaminants; Rotor And Accessories; Autoclaving - Hettich EBA 20 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

EN
16.1.2

Surface disinfection

If infectious materials penetrates into the centrifugal chamber this is to be disinfected immediately.
Ingredients of suitable disinfectants:
ethanol, n-propanol, isopropyl alcohol, glutardialdehyde, quaternary ammonium compounds.
After using disinfectants, remove the disinfectant residue by wiping with a damp cloth.
The surfaces must be dried immediately after disinfecting.
16.1.3

Removal of radioactive contaminants

The agent must be specifically labelled as being an agent for removing radioactive contaminants.
Ingredients of suitable agents for removing radioactive contaminants:
anionic tensides, non-ionic tensides, polyhydrated ethanol.
After removing the radioactive contaminants, remove the agent residue by wiping with a damp cloth.
The surfaces must be dried directly after removing the radioactive contaminants.
16.2

Rotor and accessories

In order to prevent corrosion and material changes, the rotor and accessories must be cleaned regularly with
soap or a mild detergent and a damp cloth. Cleaning is recommended at least once a week, even better after
every usage.
If the rotor or accessories are contaminated by pathogenic or radioactive material, appropriate cleaning must be
carried out.
Ingredients of suitable disinfectants:
glutaraldehyde, propanol, ethyl hexanol, anionic tensides, corrosion inhibitors.
Ingredients of suitable agents for removing radioactive contaminants:
anionic tensides, non-ionic tensides, polyhydrated ethanol.
The rotor and accessories must be dried immediately after cleaning.
The rotor and accessories must be checked monthly for corrosion damage.
The rotor and accessories must no longer be used if they show signs of wear or corrosion.
Check the firm seating of the rotor on a weekly basis.
16.3

Autoclaving

The rotor and accessories must not be autoclaved.
16.4

Centrifuge containers

With leakiness or after the breakage of centrifuging containers broken container parts and leaked centrifugation
material are to be completely removed.
The rubber inserts as well as the plastic sleeves of the rotors are to be replaced after a glass breakage.
Remaining glass splitters cause further glass breakage!
If this concerns infectious material, a disinfection process is to be executed immediately.
26/54

Publicité

Table des Matières
loading

Table des Matières